PL5

Refill Delay

PL5


Command Group: MassFlow

Access:

Network interfaces:

RW

CANopen

SDO2B00sub05

MODbus

2508,2

PL5 - Set the delay time after refill

PROFIbus

read 42,5 write 42,133


Synopsis:
CommandResponse
GetPL5P5:<Value>
SetPL5 <Value>OK (or ERR)
Valid forMin. valueMax. value
141.185 0 999999
142.185 0 999999
143.185 0 999999
88.185 0 262143



PL5 Delay time after refill

The delay time starts at the end of a refill period or shaking process.
Delay time PL5 will be setup in 10ms per increment.

See also  LS, LI, GF, GR, GM, PL1, PL2, PL3, PL4,
And for the DAD:  VF, GD, PL6, PL7, PL8, PL9, PL10, PL11, PL12, PL13, PL14, PL15, PL16, PL17, PL18, PL19, PL20, PL21.