PD1

Pre-fill mode

PD1


Command Group: Filling

Access:

Network interfaces:

RW

CANopen

SDO2200sub01

MODbus

2300,2

PD1 - Pre-fill mode.

PROFIbus

read 6,1 write 6,129


Synopsis:
CommandResponse
GetPD1P1:<Value>
SetPD1 <Value>OK (or ERR)
Valid forMin. valueMax. value
141.184 0 63
142.184 0 63
143.184 0 63
151.184 0 63
151.186 0 63
152.184 0 63
64.184 0 63
78.184 0 3
88.184 0 15



PD1 Pre filling mode. Select a prefill mode by adding the values below:

Note: Second prefill only available in LDM88 from FW version 88.184.v.2.20 and later.

Example1: one prefil with only fine valve; set PD1= 2.

Example2: one prefil with both valves; set PD1= 3 (1+2).

Example3: two prefills - first with fine valve, second with both; set PD1= 14 (2+4+8)..