NT

Nomotion time

NT


Command Group: Motion

Access:

Network interfaces:

RW

CANopen

SDO2100sub0B

MODbus

2114,2

NT - Set the stabilisation time for the [in motion] band

PROFIbus

read 18,11 write 31,11


Synopsis:
CommandResponse
GetNTT<Value>
SetNT <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 0 65535
141.184 0 65535
141.185 0 65535
141.186 0 65535
142.181 0 65535
142.184 0 65535
142.185 0 65535
142.186 0 65535
143.181 0 65535
143.184 0 65535
143.185 0 65535
151.181 0 65535
151.184 0 65535
151.186 0 65535
152.181 0 65535
152.184 0 65535
162.181 0 65535
179.181 0 65535
181.181 0 65535
64.181 0 65535
64.184 0 65535
68.181 0 65535
68.183 0 65535
69.181 0 65535
72.181 0 65535
72.184 0 65535
78.183 0 65535
78.184 0 65535
88.183 0 65535
88.184 0 65535
88.185 0 65535



The motion detection facility provides a means of disabling certain functions whenever a condition of instability, or “motion”, is detected. The “no-motion”, or “stable” condition is achieved whenever the signal is steady for the period of time set by NT, during which it cannot fluctuate by more than NR increments.The stable condition activates the relevant bit of responses to “Info Status” (IS).

The following functions are disabled if motion is detected: “Calibrate Zero” (CZ) “Calibrate Gain” (CG) “Set Zero” (SZ) and “Set Tare” (ST).

NT This sets the time (in milliseconds) over which the weight signal is checked to see if it is “stable” or has “no-motion”. The weight signal has to vary by less than NR divisions over the time period NT to be considered “stable”.

See also NR,   (IS, CZ, CG, SZ, ST).