Command Group: Calibration | Access: | Network interfaces: | |
TAC protected | RW | CANopen | SDO2300sub02 |
| | MODbus | 2202,2 |
AZ - Set the absolute zero point | | PROFIbus | read 3,2 write 4,2 |
Synopsis: |
| Command | Response |
Get | AZ | Z<Value> |
Set | AZ <Value> | OK (or ERR) |
|
Valid for | Min. value | Max. value |
141.181 |
-33000 |
33000 |
141.184 |
-33000 |
33000 |
141.185 |
-33000 |
33000 |
141.186 |
-33000 |
33000 |
142.181 |
-33000 |
33000 |
142.184 |
-33000 |
33000 |
142.185 |
-33000 |
33000 |
142.186 |
-33000 |
33000 |
143.181 |
-33000 |
33000 |
143.184 |
-33000 |
33000 |
143.185 |
-33000 |
33000 |
151.181 |
-33000 |
33000 |
151.184 |
-33000 |
33000 |
151.186 |
-33000 |
33000 |
152.181 |
-33000 |
33000 |
152.184 |
-33000 |
33000 |
162.181 |
-33000 |
33000 |
179.181 |
-33000 |
33000 |
181.181 |
-33000 |
33000 |
64.181 |
-33000 |
33000 |
64.184 |
-33000 |
33000 |
68.181 |
-22000 |
22000 |
72.181 |
-32000 |
32000 |
72.184 |
-32000 |
32000 |
78.183 |
-22000 |
22000 |
88.183 |
-22000 |
22000 |
88.184 |
-22000 |
22000 |
88.185 |
-22000 |
22000 |
|
AZ The command AZ is used as reference point for all weight calculations and will setup in mV/V. Permitted values are ± 32000 (= ± 3.2000 mV/V).
Factory default: 00000 @ 0.0000 mV/V input signal.
See also
CE,
CM,
CI,
DS,
DP,
CZ,
CG,
ZT,
AG,
CS,
FD,
UR,